Skip Navigation

Europace 2005 7(3):287; doi:10.1016/j.eupc.2005.02.103
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Catanzariti, D.
Right arrow Articles by Vergara, G.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Catanzariti, D.
Right arrow Articles by Vergara, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.

CP03: PERMANENT DIRECT HIS BUNDLE PACING DOESN'T INDUCE VENTRICULAR DYSSYNCHRONY AS OPPOSED TO CONVENTIONAL RIGHT VENTRICULAR APICAL DDD PACING

Domenico Catanzariti, M. Maines, A. Ferro, C. Cemin and G. Vergara

Cardiology Division – Ospedale S. Maria del Carmine – Rovereto (TN) – Italy

Beneficial effects of A-V synchrony due to DDD right ventricular apical (RVA) pacing could be neutralized by ventricular dyssynchrony(D).No data are reported about effects of direct his bundle pacing(DHP) on ventricular synchronism.

AIM: To assess the capability of DHP to prevent pacinginduced ventricular D as compared to DDD-RVA pacing in pts undergoing permanent DHP.

METHODS: 7 pts(2women,mean age 77+/–5)with normal HV were implanted for SSS(6 brady-tachy,1 LW II AVB) with a right atrium lead connected to the atrial channel of the device and a 4.1F screw-in hisian lead(Medtronic,Select Secure)connected to the ventricular channel.Two 4-polar leads through a femoral vein were used for DDD-RVA pacing. Fixation of permanent DHP lead was performed above the tricuspid valve using pacemapping and HB potentials. All pts underwent PW tissue Doppler imaging(TDI)analysis to assess inter- and intra-ventricular D. Interventricular mechanical delay(IVMD) between pulmonary and aortic systolic flow, septal to left posterior wall motion delay(SPWMD),and maximal difference between DTI systolic velocities of any 2 of the 4 left ventricle basal segments(maxmin DTI)were measured. DDD-RVA pacing was compared to DHP and to DHP+/–capture of right ventricular high inflow septum(RVHS)at the same atrial pacing rate after optimizing A-V interval.

RESULTS: Permanent DHP was safely obtained in 6 of 7 pts (only RVHS pacing in 1 pt). DHP and RVHS pacing thresholds, impedance and sensed potentials of DHP leads were 1.4+/–0.65 V and 1.7+/–0.8 V(x0.5 ms),611+/–162 ohm and 5.7+/–3.7mV respectively.In 5 of 6 pts DHP threshold was slightly inferior to RVHS capture threshold.Procedural and fluoroscopy times were 145 min (range 106-147) and 10.3+/–4.4 min respectively. Inter- and intra-ventricular D were reduced during DHP in comparison with DDD-RVA pacing: IVMD 198+/–36 ms vs 47+/–39 ms(cut-off value: 110; p<0.05); V-VD 45+/–18 vs 3+/–2.12 (cut-off value:30;p<0.05);max-min DTI 33.3+/–5.77 vs 19.2+/–12.5(cut-off value:10 ms).Similar data were obtained comparing DHP+RVHS and DDD-RVA pacing.No differences were observed between DHP and DHP+RVHS pacing.

CONCLUSION: DHP is feasible and safe. It prevents pacing-induced ventricular D as opposed to standard RVA DDD pacing. Longer follow-up is required to assess left ventricle remodelling prevention.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.